Handyscope HS3, a professional USB oscilloscope 订购编号:
点击这里发送询价请详细填写一下内容,包括产品原产国、品牌完整名称以及您所能得到的所有信息,这将有助于我们准确迅速的报价。非常感谢!
The Handyscope HS3 is a powerful 100 MS/s USB oscilloscope that can also be used as a multimeter, spectrum analyzer, data logger or protocol analyzer. This USB oscilloscope also contains a function / arbitrary waveform generator that is able to produce every signal that is needed. Together with the 2 measurement channels this oscilloscope is an excellent measurement tool to troubleshoot circuits.
The Handyscope HS3 is a versatile USB oscilloscope combined with a function generator in a small package.
Key specifications of the Handyscope HS3 are:
Oscilloscope / Spectrum analyzer / Multimeter / Data logger |
---|
8, 12, 14 and 16 bit user selectable resolution |
100 MS/s sampling |
50 MHz bandwidth |
128 KSamples memory |
0.2% DC vertical accuracy |
100 ppm time base accuracy |
Arbitrary Waveform Generator |
---|
Up to 2 MHz sine, square, triangular and arbitrary waves |
50 MS/s, 14 bit, 128 KSamples arbitrary waves |
-12 to 12 V output (24 Vpp) |
100 ppm time base accuracy |
900 µV RMS noise level |
The Handyscope HS3 is available in 5 models that distinguish in maximum sampling speed:
Order code | Max. sampling speed | |
---|---|---|
HS3-AWG-100 | 100 MS/s | |
HS3-AWG-50 | 50 MS/s | |
HS3-AWG-25 | 25 MS/s | |
HS3-AWG-10 | 10 MS/s | |
HS3-AWG-5 | 5 MS/s |
The Handyscope HS3 is delivered with:
Amount | Item |
---|---|
1 | Handyscope HS3 |
2 | Oscilloscope Probe 1:1-1:10 - HP-3060 |
1 | External power cable (USB) |
1 | CD-ROM with Multi Channel oscilloscope software and drivers |
1 | Instrument manual |
1 | Software manual |
The universal Multi Channel oscilloscope software that is standard delivered with the Handyscope HS3 turns your PC into a storage oscilloscope, a spectrum analyzer, a multimeter, a datalogger/transient recorder and a protocol analyzer. An intelligent auto setup allows the inexperienced user to perform measurements immediately. Complex mathematical operations can be performed live on the measured data. The software allows you to save instrument settings on disk. Later these setting files can be recalled, reducing the setup time of your instrument to an absolute minimum.
When one Handyscope HS3 does not offer enough input channels, the Handyscope HS3 can be coupled to one or more other instruments. This allows to make a combined instrument which will enable simultaneous measuring on all channels of all combined instruments.
Besides measuring in block mode, the Handyscope HS3 is also capable of performing continuous streaming measurements. This will create a continuous uninterrupted data stream to the computer. The data can then be displayed on the screen and/or saved to disk.
The integrated Arbitrary Waveform Generator operates fully independent from the measuring system of the Handyscope HS3. It has a 50 MS/s sampling clock, a 256K samples deep waveform memory, a 14 bit output resolution and a ± 12 V output range. The arbitrary function generator allows you to generate test signals for your measurements in any shape you like. Several standard signal shapes are available and it is possible to regenerate previously measured signals.
Connected to the fast USB 2.0 (High Speed, 480 Mbit/s), the Handyscope HS3 doesn't require an external power supply, therefore the Handyscope HS3 easily complies to "Plug in and measure".
The tables below show detailed specifications of the Handyscope HS3. Jump to acquisition specifications, function generator specifications or general specifications.
Acquisition system | |
---|---|
Number of input channels | 2 analog |
CH1, CH2 | BNC |
Type | Single ended |
Resolution | 8, 12, 14 and 16 bit user selectable |
Accuracy | 0.2 % of full scale ± 1 LSB |
Ranges | 200 mV to 80 V full scale |
Coupling | AC/DC |
Impedance | 1 MOhm / 30 pF |
Maximum input voltage (in all ranges) | 200 V (DC + AC peak < 10 kHz) |
Maximum input voltage 1:10 probe (in all ranges) | 600 V (DC + AC piek < 10 kHz) |
Bandwidth (-3dB) | DC to 50 MHz maximum |
AC coupling cut off frequency (-3dB) | 1.5 Hz |
Maximum sampling rate | Depending on model, on all channels simultaneously |
Model HS3-100 | |
8 bit | 100 MS/s |
12 bit | 50 MS/s |
14 bit | 3.125 MS/s |
16 bit | 195 kS/s |
Model HS3-50 | |
8 bit | 50 MS/s |
12 bit | 50 MS/s |
14 bit | 3.125 MS/s |
16 bit | 195 kS/s |
Model HS3-25 | |
8 bit | 25 MS/s |
12 bit | 25 MS/s |
14 bit | 3.125 MS/s |
16 bit | 195 kS/s |
Model HS3-10 | |
8 bit | 10 MS/s |
12 bit | 10 MS/s |
14 bit | 3.125 MS/s |
16 bit | 195 kS/s |
Model HS3-5 | |
8 bit | 5 MS/s |
12 bit | 5 MS/s |
14 bit | 3.125 MS/s |
16 bit | 195 kS/s |
Maximum stream rate | 10 kS/s |
Sampling clock source | |
Internal | Quartz |
Accuracy | ±0.01 % |
Stability | ±100 ppm over -40 °C to +85 °C |
Time base aging | ±5 ppm/year |
External | Extension connector |
Memory | 128 KiSamples per channel |
Trigger | |
System | Digital, 2 levels |
Source | CH1, CH2, AND, OR, digital external, generator start, generator new period, generator stop |
Trigger modes | Rising edge, falling edge, inside window, outside window |
Level adjustment | 0 to 100 % of full scale |
Hysteresis adjustment | 0 to 100 % of full scale |
Resolution | 0.025 % (12 bits) |
Pre trigger | 0 to 128 KiSamples, 1 sample resolution |
Post trigger | 0 to 128 KiSamples, 1 sample resolution |
Digital external trigger | |
Input | Extension connector |
Range | 0 to 3.3 V (5 V max) |
Coupling | DC |
Arbitrary Waveform Generator (independent from acquisition system) | |
Waveforms | Sine, triangle, square, DC, noise and arbitrary |
Number of output channels | 1 analog, BNC |
DAC resolution | 14 bit @ 50 MS/s |
Output range | -12 V to 12 V |
Amplitude | |
Range | 0.1 V, 0.9 V, 12 V |
Resolution | 13 bit |
Accuracy | 0.4 % |
DC offset | |
Range | -12 V to 12 V |
Resolution | 13 bit |
Coupling | DC |
Impedance | 50 Ohm |
Bandwidth | DC to 2 MHz |
Noise level |
0 to ±0.1 V range : 900 µV RMS 0 to ±0.9 V range : 1.3 mV RMS 0 to ±12 V range : 1.5 mV RMS |
System characteristics | |
System | DDS |
Memory | 256 KiSamples |
Maximum Sampling rate | 50 MHz |
Sampling source | internal |
Accuracy | ±0.01 % |
Stability | ±100 ppm over -40 °C to +85 °C |
Interface | |
Interface | USB 2.0 High Speed (480 Mbit/s); (USB 1.1 Full Speed (12 Mbit/s) and USB 3.0 compatible) |
Power Requirements | |
Power from USB port | 500 mA max (2.5 W max) |
Power via external power input / extension connector | 1500 mA max (7.5 W max) |
Minimum voltage | 4.5 VDC |
Maximum voltage (SN# <12941) | 6 VDC |
Maximum voltage (SN# >12941) | 12 VDC |
Physical | |
Instrument height | 25 mm (1 inch) |
Instrument length | 170 mm (6.7 inch) |
Instrument width | 140 mm (5.2 inch) |
Cord length | 1.8 m (70 inch) |
Weight | 480 g (17 ounce) |
I/O connectors | |
Channel 1, 2 | BNC |
Generator | BNC |
USB | fixed cable with USB 2.0 and USB 1.1 type A connector |
Extension connector | D-sub 25 pins female |
System requirements | |
PC I/O connection | USB 2.0 High Speed (480 Mbit/s); (USB 1.1 Full Speed (12 Mbit/s) and USB 3.0 compatible) |
Operating system | Windows 98/ME/2000/XP/Vista/7/8/10 |
Operating environment | |
Ambient temperature | 0 to 55°C |
Relative humidity | 10 to 90%, non condensing |
Storage environment | |
Ambient temperature | -20 to 70°C |
Relative humidity | 5 to 95%, non condensing |
Certifications and Compliances | |
CE mark compliance | yes |
RoHS | yes |
Package | |
Instrument | Handyscope HS3 |
Oscilloscope probes | 2 x Oscilloscope Probe 1:1-1:10 - HP-3060 |
Accessories | external power cable for second USB port |
Software | for Windows 98/2000/ME/XP/Vista/7/8/10 on CD |
Drivers | for Windows 98/2000/ME/XP/Vista/7/8/10 on CD |
Manuals |
instrument manual and software user's manuals color printed and digital on CD |
Options | |
Optionally available | 10 MHz generator output |
Optionally available | Single channel Handyscope HS3 without generator, but with 512K samples memory |
Several accessories are available for use with the Handyscope HS3. The most common related products, such as oscilloscope probes and current clamps are listed here. Refer to the products page to find all TiePie products.
The Handyscope HS3 is available in 5 models that distinguish in maximum sampling speed:
Order code | Max. sampling speed |
---|---|
EJH3-AWG-100 | 100 MS/s |
EJH3-AWG-50 | 50 MS/s |
EJH3-AWG-25 | 25 MS/s |
EJH3-AWG-10 | 10 MS/s |
EJH3-AWG-5 | 5 MS/s |