1007 DC Millivolt Potentiometer & Calibrator 订购编号:
点击这里发送询价请详细填写一下内容,包括产品原产国、品牌完整名称以及您所能得到的所有信息,这将有助于我们准确迅速的报价。非常感谢!
The 1007 is a handheld calibrator that can be used for potentiometric voltage measurement in addition to its operation as a millivolt source.
The null zero and sensitivity are adjustable via front panel controls. Maximum sensitivity enables null balance to resolve 3 microvolt.
Three output ranges are provided to give adjustable output values from 1μV to 1V with a basic 0.02% accuracy. For signal injection, the operator needs to switch on, check the battery condition, select the range, and set the required voltage using the thumbwheel switches. The 1007 uses a precision reference diode and low temperature coefficient resistors to give a highly stable output.
Power is provided by 6 AA batteries. Battery life is several months, depending on usage. The battery condition is monitored by an indicator mounted on the top of the unit. The 1007 has up to 20mA drive current and is short circuit and overload protected. An off/normal/reverse output polarity switch is provided.
Safety Terminals
Fitted as standard and fully compatible with 4mm shrouded plugs, as well as standard plugs, bare wires, and spade terminals.
Added Protection
The 1007 comes fitted with an ergonomic rubber cover providing increased protection and durability. It has a textured grip for comfortable handling and openings at the top and bottom to allow access to the battery meter and a position to place labels if required. It is easy to remove if the user prefers a stand-alone unit or to house the 1007 in the 9027 carry case.
Applications
Suitable for calibration and simulation of thermocouples. Accurate voltages equivalent to the output from a thermocouple can easily be set on the 1007, enabling fast calibration of temperature measuring equipment.
Alternatively, the 1007 can measure thermocouples output by operating as a potentiometer. Other applications include chart recorder calibration, A/D converter and DMM calibration, and use as a stable voltage for backing off DC offsets.
The 1007 is an easy to use test instrument designed for accuracy and reliability. As part of the Time Electronics portable calibration equipment range it is ideal for use in the lab or field.